Shown with 45mm Shutter
See the MicroLine page for details and alternate configurations.
The back-illuminated CCD in the ML4710 has exceptionally high quantum efficiency. Larger pixels provide deep dynamic range and gather more light for improved signal-to-noise. Available with midband, broadband, or UV coating on the sensor.
Link to ML4710 Specification Sheet in pdf format.
Link to C-mount Version of ML4710 in pdf format.
Sensor Type | Back Illuminated CCD |
Sensor | Teledyne e2v CCD47-10 |
Active Pixels | 1024 x 1024 |
Pixel Size (microns) | 13 x 13 µm |
Imaging Area (Diagonal) | 13.3 X 13.3 mm (18.8 mm) |
Full Well Capacity | 100000 electrons |
Typical Readout Noise | 10 e- @ 700 kHz; 17 e- @ 2 MHz |
Typical Gain | 1.3 e-/ADU |
Dynamic Range | 79.7 dB |
Anti-Blooming | None |
Cooling Method | Air (Optional Liquid) |
Max. Cooling (Air) | 60°C below ambient |
Temperature Stability | 0.1°C |
Dark Current (typical) | 0.1 eps at -40C |
Interface | USB 2.0 |
Digitization Clock | 700 kHz and 2 MHz |
Data Bit Depth | 16 bit |
Non-Linearity | <1% |
Channels | 1 |
Shutter | 65 mm |
Optional Mount | Nikon F-mount |
Subarray Readout | Standard |
External Trigger In/Out | Standard |
SDK / Software | USB2 / FLIGrab |
Weight | 2.8 lbs (1.2 kg) |
Environment | -30°C to 45°C | 10% - 90% Relative Humidity |
Power | 12V (100-240V AC to 12V DC power supply included). With TEC off: <1A. TEC at 100%: 4.4A. (Optional Electromechanical Shutter open: 4A pulse for 100msec. Shutter held open, add 0.22A.) |